I'm trying to read weight from Sartorius Weighing Scale model No BS2202S using the following code in C#.net 2.0 on a Windows XP machine:
public string readWeight()
{
string lastError = "";
string weightData = "";
SerialPort port = new SerialPort();
port.PortName = "COM1";
port.BaudRate = 9600;
port.Parity = Parity.Even;
port.DataBits = 7;
port.StopBits = StopBits.One;
port.Handshake = Handshake.RequestToSend;
try {
port.Open();
weightData = port.ReadExisting();
if(weightData == null || weightData.Length == 0) {
lastError = "Unable to开发者_StackOverflow read weight. The data returned form weighing machine is empty or null.";
return lastError;
}
}
catch(TimeoutException) {
lastError = "Operation timed out while reading weight";
return lastError;
}
catch(Exception ex) {
lastError = "The following exception occurred while reading data." + Environment.NewLine + ex.Message;
return lastError;
}
finally {
if(port.IsOpen == true) {
port.Close();
port.Dispose();
}
}
return weightData;
}
I'm able to read the weight using Hyperterminal application (supplied with Windows XP) with the same serial port parameters given above for opening the port. But from the above code snippet, I can open the port and each time it is returning empty data.
I tried opening port using the code given this Stack Overflow thread, still it returns empty data. Kindly assist me.I know this is probably old now ... but for future reference ...
Look at the handshaking. There is both hardware handshaking and software handshaking. Your problem could be either - so you need to try both.
For hardware handshaking you can try:
mySerialPort.DtrEnable = True
mySerialPort.RtsEnable = True
Note that
mySerialPort.Handshake = Handshake.RequestToSend
I do not think sets the DTR line which some serial devices might require
Software handshaking is also known as XON/XOFF and can be set with
mySerialPort.Handshake = Handshake.XOnXOff
OR
mySerialPort.Handshake = Handshake.RequestToSendXOnXOff
You may still need to enable DTR
When all else fails - dont forget to check all of these combinations of handshaking.
Since someone else will probably have trouble with this in the future, hand shaking is a selectable option.
In most of the balances you will see the options for Software, Hardware 2 char, Hardware 1 char. The default setting for the Sartorius balances is Hardware 2 Char. I usually recommend changing to Software.
Also if it stops working all together it can often be fixed by defaulting the unit using the 9 1 1 parameter. And then resetting the communication settings.
An example of how to change the settings can be found on the manual on this page: